Z-1762-2022Class II
NanoSight LM10 fitted with LM14 module. A laboratory instrument for determination of particle characteristics...
Malvern Panalytical Ltd / initiated 2022-02-28 / Open, Classified
MALVERN PANALYTICAL LTD began a recall of NanoSight LM10 fitted with LM14 module. A laboratory instrument for determination of particle characteristics (e.g. size) on . FDA classified it as Class II and gives the reason as "Product found to not comply to Class 1 Laser safety requirement". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1762-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Malvern Panalytical Ltd
- Product
- NanoSight LM10 fitted with LM14 module. A laboratory instrument for determination of particle characteristics (e.g. size).
- Product code
- REN Particle-Size Measuring Instrument
- Reason for recall
- Product found to not comply to Class 1 Laser safety requirement.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.
- Quantity in commerce
- 76 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- CDRH accession # 0510530-002