Z-0082-2023Class II

Prismaflex M Set: M150 (product code 109990). For use in providing continuous fluid management and renal...

Baxter Healthcare Corporation / initiated 2022-09-14 / Open, Classified / root cause: labeling and UDI

Baxter Healthcare Corporation began a recall of Prismaflex M Set: M150 (product code 109990). For use in providing continuous fluid management and renal replacement therapies on . FDA classified it as Class II and gives the reason as "Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0082-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Prismaflex M Set: M150 (product code 109990). For use in providing continuous fluid management and renal replacement therapies.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Baxter sent an Urgent Medical Device Correction notice to its direct consignees by USPS on 09/14/2022. The notice explained the Estonian translation issue and requested that the following actions be taken by consumers: 1. Use of the Prismaflex HF20 set should be restricted to patients with a body weight greater than 8kg (18lb). Use of the Prismaflex M60 and ST60 sets should be restricted to patients with a body weight greater than 11kg (24lb). Use of the Prismaflex M100, ST100, M150, ST150, HF1000 and HF1400 sets should be restricted to patients with a body weight greater than 30kg (66lb). 2. Customers who are not referring to the Estonian IFU should continue to follow the IFU instructions in their language. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com/. 4. If you purchased this product from a distributor, please note that responding via the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal.
Quantity in commerce
334,932 devices
Distribution
US Nationwide
Product codes and lots
UDI 07332414090005; All Lot Codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0082-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0081-2023Class IIOngoingVoluntary: Firm initiated
Z-0082-2023Class IIOngoingVoluntary: Firm initiated
Z-0083-2023Class IIOngoingVoluntary: Firm initiated
Z-0084-2023Class IIOngoingVoluntary: Firm initiated
Z-0085-2023Class IIOngoingVoluntary: Firm initiated