Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0008-2023Class II

Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Synthes (USA) Products LLC / initiated 2022-08-30 / Open, Classified / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001 on . FDA classified it as Class II and gives the reason as "Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0008-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes (USA) Products LLC
Product
Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Product code
JDR Staple, Fixation, Bone
Reason for recall
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
DePuy Synthes issued Urgent Medical Device Recall (Removal) letter on 8/30/22. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the subject product and quarantine. DO NOT USE THE SUBJECT PRODUCT. 2. Contact your DePuy Synthes Sales Consultant or call the customer support services at 1-800-255-2500 to coordinate the return/credits of the subject product. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to [email protected] within 3 business days of receipt of this notification. Please include in the email subject: FA 2154225: 2022 Hammertoe CCI Kit. " Please complete the attached Business Response Form even if you do not have the subject product on hand. 4. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 5. If any of the subject product has been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records. Inquiries please do not hesitate to contact your DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.
Quantity in commerce
191 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0008-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0008-2023Class IIOngoingVoluntary: Firm initiated