Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room...
Baxter Healthcare Corporation began a recall of Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table on . FDA classified it as Class II and gives the reason as "the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0175-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table
- Product code
- GDC Table, Operating-Room, Electrical
- Reason for recall
- the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- FDA root cause
- Other
- Action
- The firm issued an Urgent Medical Device Correction notice to their consignee on 09/20/2022 by email. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot number listed on page 1 and stop any use of the product immediately. 2. Hillrom will contact you to perform a correction of your impacted device as soon as possible." The firm issued an Urgent Medical Device Recall notice to their consignees on 09/27/2022 by FedEx 2nd day delivery. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot numbers listed in the above table and replace with the enclosed device that accompanies this notification. 2. Please return the impacted unit per RMA instructions attached within 10 business days. Actions to be taken by Distributor: 1. Please forward this notification to your end users along with replacement parts accompanying this notification. 2. Please request return of the impacted devices from your end users and return the devices to Hillrom, per RMA instructions attached." Both notices provided the following contact information: "Contact reference person If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at 800-445-3720 or [email protected]"
- Quantity in commerce
- 2 units
- Distribution
- MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
- Product codes and lots
- a) UDI/DI 00615521003012, Serial Number 122190000035, Lot Number 122220000528 b) UDI/DI 00615521003036, Serial Number 122230000655, Lot Number 122180001460
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0175-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0176-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0177-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0178-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0179-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0180-2023 | Class II | Terminated | Voluntary: Firm initiated |