Z-0114-2023Class II
Medicina Reusable Oral syringe 2.5ml Code: OTH25
Medicina Ltd / initiated 2022-07-04 / Open, Classified / root cause: manufacturing process
Medicina Uk Ltd began a recall of Medicina Reusable Oral syringe 2.5ml Code: OTH25 on . FDA classified it as Class II and gives the reason as "Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0114-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medicina Ltd
- Product
- Medicina Reusable Oral syringe 2.5ml Code: OTH25
- Product code
- KYW Container, Liquid Medication, Graduated
- Reason for recall
- Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medicina Ltd issued Field Safety Corrective Action notice to the US Distributor 07/04/2022 via email. The US Distributor is conducting recall to the user level letter issued the week of July 25th via United States Postal Service. . Letter states reason for recall, health risk and action to take: Action to be taken - Please carefully read this Field Safety Notice and return the Acknowledgement Form - Please do not use any of the batches of product listed on this notice - If you have product to return, please complete the information on the Acknowledgement Form and our team will arrange the return for you - Please distribute throughout your organisations and to relevant personnel - If you have any questions or queries in relation to this notice, please contact us using the information below Contact information regarding FSN Shaista Patel QA/RA Administrator Unit 2, Rivington View Business Park, Station Road, Blackrod, Bolton BL6 5BN Email: [email protected] Tel: 01204 695050
- Quantity in commerce
- 20 boxes (2000 syringes)
- Distribution
- Nationwide
- Product codes and lots
- UDI-DI: (01)05056115402876 Batch number: HM18042, HM21019, HM21054
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0107-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0108-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0109-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0110-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0111-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0112-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0113-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0114-2023 | Class II | Ongoing | Voluntary: Firm initiated |