G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G)...

Zimmer GmbH / initiated 2022-08-25 / Open, Classified / root cause: packaging

Zimmer, Inc. began a recall of G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty on . FDA classified it as Class II and gives the reason as "The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0007-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
Product code
OQI Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented
Reason for recall
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
On August 25, 2022, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were instructed to quarantine all affected product. Zimmer Biomet sales representatives will remove the affected product from customer facilities. Customers should ensure that affected personnel are aware of the recall, and also notify any recipients if the product has been further distributed. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
56 US
Distribution
Domestic distribution US nationwide.
Product codes and lots
Item No. 010000663 UDI-DI: 00880304524224 Lots: 7286477 7289929 7296576 7296633 7301697 Exp. 21-Jul-32 Item No. 010000667 UDI-DI: 00880304524262 Lot: 7253325 Exp. 13-Jun-32

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0007-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0007-2023Class IIOngoingVoluntary: Firm initiated