Z-0138-2023Class II

CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B...

Merit Medical Systems, Inc. / initiated 2022-08-03 / Open, Classified / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB... on . FDA classified it as Class II and gives the reason as "Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0138-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Merit Medical Systems, Inc.
Product
CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB, CORCA1615S/CNB, CORCA1810SB/CNB, CORCA1815SB/CNB, CORCA1820SB/CNB, CORCA2015SB/CNB, CORCA2020SB/CNB
Product code
KNW Instrument, Biopsy
Reason for recall
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
On 8/3/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify affected devices, quarantine, discontinue use and distribution, and return to the recalling firm. 2) If product was further distributed, share the recall notice with these customers/users. 3) Complete and return the customer response form. Customers with questions can contact Customer Service at 801-208-4381, 6AM-6PM, Mon-Fri.
Quantity in commerce
2894
Distribution
Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France, Canada, Mexico, Puerto Rico, Chile, Reunion, Belgium, Switzerland, Japan, South Africa, Qatar, Russian Federation, Hungary, Thailand, Namibia, Colombia, New Zealand, Netherlands, Saudi Arabia, Costa Rica, Oman, Cyprus, Slovenia, Croatia, China.
Product codes and lots
Catalog Code-Lot (OUS Distribution Only): CORA1410/B- I2206909, I2318475, I2291980, I2286704; CORA1410S/B- I2206888; CORA1415/B- I2213361; CORA1615S/B- I2114518, H1517631, H1715532, H1756745, H1767115, H1771762, H1776067, H1784861, H1800979, H1808121, H1825693, I1781878, I1818817, I1837234, I1843704, I1855418, I1866359, I1952354, I1958525, I1971876, I1990075, I1998862, I2043630, I2070249, I2075810; CORA2010S/B- I2010414; CORCA1410S/B- I2259093; CORCA1410S/CNB- I2305358; CORCA1415S/B- I2210950; CORCA1610/CNB- I2291983; CORCA1615S/CNB- I2330683; CORCA1810SB/CNB- I2305359; CORCA1815SB/CNB- I2170254, I2205164; CORCA1820SB/CNB- I2325912; CORCA2015SB/CNB- I2318903, I2325926; CORCA2020SB/CNB- I2320880

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0138-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0137-2023Class IIOngoingVoluntary: Firm initiated
Z-0138-2023Class IIOngoingVoluntary: Firm initiated