Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay
Siemens Healthcare Diagnostics, Inc. began a recall of Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay on . FDA classified it as Class II and gives the reason as "Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0104-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay
- Product code
- NDW Assay, Porphyrin, Spectrophotometry, Lithium
- Reason for recall
- Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.
- FDA root cause
- Under Investigation by firm
- Action
- On September 19, 2022, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers should take the following actions: -Please review this letter with your Medical Director. -Discontinue use of and discard the affected product. Siemens is temporarily discontinuing production and distribution of the Atellica CH lithium reagents until the issue is resolved. In the interim, Siemens is recommending that customers transition to an alternate lithium assay.
- Quantity in commerce
- 1342 US; 2266 OUS
- Distribution
- Worldwide distribution - US Nationwide.
- Product codes and lots
- UDI-DI: 00630414287935 All lots impacted: Lots 110238 exp 10/18/22 110302 exp 12/13/22 120017 exp 4/18/23
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0104-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0105-2023 | Class II | Ongoing | Voluntary: Firm initiated |