Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0099-2023Class II

SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512

Synthes (USA) Products LLC / initiated 2022-09-27 / Open, Classified / root cause: labeling and UDI

Synthes (USA) Products LLC began a recall of SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512 on . FDA classified it as Class II and gives the reason as "The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0099-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes (USA) Products LLC
Product
SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512
Product code
JDR Staple, Fixation, Bone
Reason for recall
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
On September 27, 2022, the firm notified affected customers via urgent medical Device Recall letters. Customers were instructed to examine their inventory to determine if they have affected product and to quarantine them immediately - Customers should not use the recalled products. Customers should contact their DePuy Synthes Sales Consultant or call the customer support services at [email protected] to coordinate the return/credits of the subject products.
Quantity in commerce
94 US; 30 OUS
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI (GTIN): 00810633020166 Lot/Expiry MSE220445 7/1/2027 MSE210184 8/1/2026

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0099-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0098-2023Class IIOngoingVoluntary: Firm initiated
Z-0099-2023Class IIOngoingVoluntary: Firm initiated
Z-0100-2023Class IIOngoingVoluntary: Firm initiated
Z-0101-2023Class IIOngoingVoluntary: Firm initiated