WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System
MICROVENTION INC. began a recall of WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System on . FDA classified it as Class II and gives the reason as "Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0170-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
- Product code
- OPR Intrasaccular Flow Disruption Device
- Reason for recall
- Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 09/21/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify, quarantine, and return all impacted devices. 2) Complete and return Customer Acknowledgment and Device Reconciliation form the recall instructions provided in the letter and have shared this notification with all device users within my facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred. Complete and return the acknowledgement and receipt form. Customers with additional questions can call the recalling firm at 949-501-0355, Monday - Friday PST 8:00 - 5:00 pm or email [email protected]
- Quantity in commerce
- 2018
- Distribution
- Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.
- Product codes and lots
- REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345; Distributed in China: WDC/00842429113684/0000078840, 0000086571, 0000088347, 0000088349, 0000136477, 0000181460, 0000191147; Distribution in Japan: MV-WDC2/04987892122231/0000142172; 0000166445; 0000210109; 0000245337
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0170-2023 | Class II | Ongoing | Voluntary: Firm initiated |