Z-0253-2023Class II

3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020, general surgery drape

3M Healthcare Business / initiated 2022-10-06 / Open, Classified

3M Healthcare Business began a recall of 3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020, general surgery drape on . FDA classified it as Class II and gives the reason as "During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0253-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
3M Healthcare Business
Product
3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020, general surgery drape
Product code
PUI Drape, Surgical, Exempt
Reason for recall
During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.
FDA root cause
Component change control
Action
The firm began disseminating an URGENT Product Recall notice their consignees by email or letter beginning 10/06/2022. The notice explained the issue and requested the following: Distributors: "Actions Requested: 1. Immediately discontinue distribution of affected models and lots of 3M" Steri-Drape" Surgical Drapes. Reference Attachment 1 for list of models and lots. 2. Determine if any stock of the affected models and lots remain in inventory. 3. Complete the Distributor Acknowledgement Form with the following: " Acknowledge that you have read and understood this letter whether or not you have any 3M" Steri-Drape" Surgical Drapes in stock at your facility; and " Indicate if you have any product in stock and dispose of this material per facility policy. " Email the completed form to [email protected]. 4. Notify any of your customers who have purchased this product using the Customer Letter and Customer Acknowledgement Form provided. 5. To request credit, contact 3M Health Care Customer Helpline at 1-800-228-3957, option 7 or by email to [email protected]." Direct accounts: "Actions Requested: 1. Immediately cease use of affected models and lots of 3M" Steri-Drape" Surgical Drapes. Reference Attachment 1 for list of models and lots. 2. Determine if any stock of the affected models and lots remain in inventory. NOTE: if you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 3. Complete the Customer Acknowledgement Form with the following: " Acknowledge that you have read and understood this letter whether or not you have any 3M" Steri-Drape" Surgical Drapes in stock at your facility; and " Indicate if you have any product in stock and dispose of this material per facility policy. " Email the completed form to [email protected]. 4. To request credit, contact 3M Health Care Customer Helpline at 1-800-228-3957, option 7 or by email t
Quantity in commerce
769,810 units
Distribution
Worldwide distribution.
Product codes and lots
UDI/DI (01)30707387018772, Lot numbers: 33H9T5, 33HA9D, 33HC7R, 33HCJN, 33HCPL, 33HD6C, 33HDYW, 33HEDM, 33HELH, 33HF3X, 33HFXP, 33HH8C, 33HHKC, 33HJ3K, 33HK5H, 33HKJH, 33HKPA, 33HLKE, 33HLWT, 33HNJ3, 33HNXX, 33HY4Y, 33HYCP, 33J3LW, 33J43T, 33J4E7, 33J4M7, 33J5CP, 33J798, 33J7KC, 33J7YJ, 33J8A7, 33J97T, 33J9JR, 33J9RC, 33JA5D, 33JAED, 33JC7L, 33JCHH, 33JCPY, 33JD5A, 33JE49, 33JEEM, 33JEP8, 33JFAJ, 33JFJX, 33JPDM, 33JPPC, 33JR3D, 33JRCR, 33JRNA, 33JTFX, 33JTNY, 33JW4L, 33JWD4, 33JWLR, 33JXEM, 33JXRJ, 33JY7J, 33JYFL, 33JYMH, 33K3AE, 33K3N7, 33K6CJ, 33K7CL, 33K7NF, 33K9A3, 33K9JW, 33K9R5, 33K9YW, 33KAXL, 33KC8P, 33KCJT, 33KCX3, 33KE37, 33KHXA, 33KJAA, 33KJL5, 33KJWK, 33KKMY, 33KL5W, 33KLMN, 33KM4C, 33KN54, 33KNAX, 33KR5C, 33KRDE, 33KRR5, 33KTNC, 33KW4K, 33KWEX, 33KWT5, 33KXRH, 33KYA3, 33KYKA, 33KYWT, 33L3YR, 33L6EW, 33L6M5, 33L7RH

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0253-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0247-2023Class IIOngoingVoluntary: Firm initiated
Z-0248-2023Class IIOngoingVoluntary: Firm initiated
Z-0249-2023Class IIOngoingVoluntary: Firm initiated
Z-0250-2023Class IIOngoingVoluntary: Firm initiated
Z-0251-2023Class IIOngoingVoluntary: Firm initiated
Z-0252-2023Class IIOngoingVoluntary: Firm initiated
Z-0253-2023Class IIOngoingVoluntary: Firm initiated
Z-0254-2023Class IIOngoingVoluntary: Firm initiated
Z-0255-2023Class IIOngoingVoluntary: Firm initiated
Z-0256-2023Class IIOngoingVoluntary: Firm initiated
Z-0257-2023Class IIOngoingVoluntary: Firm initiated
Z-0258-2023Class IIOngoingVoluntary: Firm initiated
Z-0259-2023Class IIOngoingVoluntary: Firm initiated
Z-0260-2023Class IIOngoingVoluntary: Firm initiated
Z-0261-2023Class IIOngoingVoluntary: Firm initiated
Z-0262-2023Class IIOngoingVoluntary: Firm initiated
Z-0263-2023Class IIOngoingVoluntary: Firm initiated
Z-0264-2023Class IIOngoingVoluntary: Firm initiated