Z-0365-2023Class II
FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
Full Vision Inc / initiated 2022-08-31 / Open, Classified
Full Vision Inc began a recall of FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927 on . FDA classified it as Class II and gives the reason as "This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0365-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Full Vision Inc
- Product
- FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
- Product code
- IOL Treadmill, Powered
- Reason for recall
- This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
- FDA root cause
- Other
- Action
- The firm began sending URGENT MEDICAL DEVICE CORRECTION notices its consignees by email beginning 09/13/2022. The distributors were provided with a letter to distribute to their customers (end users) The notice explained the issue, risk to health, and requested the following actions be taken: "Actions to be taken by the Customer/User: A simple test can be performed to determine if the unit has the incorrect component installed on the PCB Assembly. The user should turn the device on. Then cycle the power with the power switch by turning it off, waiting 5 seconds, then turn it back on. Start the belt at 2 mph and pull the Magnetic Tether. If the assembly has the incorrect component the device should fail to stop. If the assembly has the correct component the device should stop. The assemblies that fail to stop should be replaced with the c-clip style safety tethers, provided by Full Vision to the distributors and end users. To continue using the treadmill with the Magnetic Tether assemblies until they can be replaced, you can prevent this condition from occurring by cycling the power for 30 seconds before turning the device back on. Product Correction: Full Vision has scrapped all the non-conforming components (Magnetic Safety Tether and PCB Board Assembly) from the inventory. All nonconforming components will be replaced."
- Quantity in commerce
- 38 units
- Distribution
- Wi, GA, NY, Mexico
- Product codes and lots
- UDI/DI 00860176000613, Serial Numbers: FVDC-7630, FVDC-7631, FVDC-7633, FVDC-7634, FVDC-7635, FVDC-7636, FVDC-7637, FVDC-7638, FVDC-7639, FVDC-7647, FVDC-7648, FVDC-7649, FVDC-7650, FVDC-7651, FVDC-7652, FVDC-7653, FVDC-7654, FVDC-7655, FVDC-7656, FVDC-7657, FVDC-7658, FVDC-7659, FVDC-7660, FVDC-7661, FVDC-7662, FVDC-7663, FVDC-7664, FVDC-7665, FVDC-7707, FVDC-7708, FVDC-7709, FVDC-7710, FVDC-7711, FVDC-7712, FVDC-7713, FVDC-7714, FVDC-7715, FVDC-7716
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0364-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0365-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0366-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0367-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0368-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0369-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0370-2023 | Class II | Ongoing | Voluntary: Firm initiated |