Z-0239-2023Class II

Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

Maquet Cardiovascular Us Sales, LLC / initiated 2022-10-14 / Open, Classified / root cause: manufacturing process

Getinge Usa Sales Inc began a recall of Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN on . FDA classified it as Class II and gives the reason as "The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0239-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
On October 14, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were provided with extra precautions to take in order to continue use of the device. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (select option 2), Monday through Friday, 8AM - 6PM ET.
Quantity in commerce
5 US; 1008 OUS
Distribution
Worldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN. The countries of Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Bulgaria (BG), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Estonia (EE), Finland (FI), France (FR), Germany (DE), Greece (GR), Hungary (HU), Indonesia (ID), Israel (IL), Italy (IT), Japan (JP), Jordan (JO), Kuwait (KW), Lithuania (LT), Luxembourg (LU), Mexico (MX), Montenegro (ME), Morocco (MA), Netherlands (NL), Norway (NO), Pakistan (PK), Peru (PE), Poland (PL), Portugal (PT), Romania (RO), Russia (RU), Saudi Arabia (SA), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), Tunisia (TN), Turkey (TR), United Arab Emirates (AE), United Kingdom (GB), Yemen (YE).
Product codes and lots
No UDI - product discontinued in 2008. Part Number 1140.20F0: Serial Nos. 3, 8, 9, 10 Part Number 1140,20AN: Serial No. 851

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0239-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0239-2023Class IIOngoingVoluntary: Firm initiated