Z-0271-2023Class II

K-ASSAY IgA Immunoturbidimetric Assay, REF: KAI-013

Kamiya Biomedical Company, LLC / initiated 2022-09-28 / Open, Classified

Kamiya Biomedical Company, LLC began a recall of K-ASSAY IgA Immunoturbidimetric Assay, REF: KAI-013 on . FDA classified it as Class II and gives the reason as "IgA Reagent may start showing cloudiness over time, which can affect assay performance". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0271-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Kamiya Biomedical Company, LLC
Product
K-ASSAY IgA Immunoturbidimetric Assay, REF: KAI-013
Product code
CFF Immunoelectrophoretic, Immunoglobulins, (G, A, M)
Reason for recall
IgA Reagent may start showing cloudiness over time, which can affect assay performance.
FDA root cause
Under Investigation by firm
Action
On 9/28/22, recall notices were sent to customers informing them to stop using affected kit lots and to return complete and partially used kits to the recalling firm. Further, if opaque turbidity is observed, or the controls of the affected assay are not within range, do not report the test results. Distributors are asked to contact their customers and to inform them of the recall. Alternately, distributors can send the recalling firm their customer list so the recalling firm can notify the customers directly. Complete and return the customer acknowledgement. The following firm contacts were provided: (206) 575-8068 and [email protected]
Quantity in commerce
165
Distribution
Worldwide - US Nationwide distribution in the states of FL, NC, TX, TN, PA, CA, NY, IL, MA and the country of Czech Republic.
Product codes and lots
UDI-DI: 00816426020092, Lot/ Expiration Date: H180/ 2023-01-31, K181/ 2023-03-31, N182/ 2023-06-30, D283/ 2023-09-30

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0271-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0271-2023Class IIOngoingVoluntary: Firm initiated