Z-0235-2023Class II
Atrium Ocean Wet Suction Water Seal Chest Drain-
Atrium Medical Corporation / initiated 2022-10-18 / Open, Classified
Atrium Medical Corporation began a recall of Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000 on . FDA classified it as Class II and gives the reason as "Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0235-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Atrium Medical Corporation
- Product
- Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000
- Product code
- KDQ Bottle, Collection, Vacuum
- Reason for recall
- Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.
- FDA root cause
- Storage
- Action
- Atrium/ Getinge issued Urgent Medical Device Removal on 10/18/21 via email with Delivery and Read receipt. Letter states reason for recall, health risk and action to take: Our records indicate that McKesson has received the Atrium Ocean Wet Suction Water Seal Chest Drains from the lot listed on Page 1. " Please examine your inventory and confirm that you have no remaining Atrium Ocean Wet Suction Water Seal Chest Drain 2002-000 with the REF/LOT number listed in this notice. Please return any product that may be affected under this Field Action. " Please respond to this email using the response form on page 3 to acknowledge that you have received this notification. If you have any questions, please contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Quantity in commerce
- 24 units (4cases/6)
- Distribution
- US Nationwide distribution in the state of VA.
- Product codes and lots
- UDI-DI: 20650862100017 - case / 00650862100013- each Lot Number: 472326, 473226
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0235-2023 | Class II | Ongoing | Voluntary: Firm initiated |