Z-0222-2023Class II

X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark...

X-NAV Technologies, LLC / initiated 2022-09-28 / Open, Classified / root cause: manufacturing process

X-NAV Technologies, LLC began a recall of X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673 on . FDA classified it as Class II and gives the reason as "Product missing one of two welds connecting the tip to the shaft, if the remaining weld breaks, it is possible that the tip could fall out during X-Mark registration and fall into the patient s mouth. In this case, the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0222-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
X-NAV Technologies, LLC
Product
X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673
Product code
QRY Dental Navigation System
Reason for recall
Product missing one of two welds connecting the tip to the shaft, if the remaining weld breaks, it is possible that the tip could fall out during X-Mark registration and fall into the patient s mouth. In this case, the patient may aspirate or swallow the probe tip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
X-Nav issued letters dated 9/30/2022 via First Class Mail to potentially affected users on 10/3/22. Letter states reason for recall, health risk and action to take: The customers were instructed to cease use of the device until you have performed the inspection described on the second page of this letter. Review the pictures on the next page and compare them to the Probe Tool in your possession. After review, contact Customer Service at the contact options listed at the bottom of this page so that we may update our records with your findings. Report any of the below options. 1. PROBE TOOL IS GOOD. If your Probe Tool matches the good Probe Tool pictures above, it may be used as-is. 2. PROBE TOOL IS BROKEN. If your Probe Tool looks like either of the broken Probe Tool pictures above, we will send you a shipping label to return your existing Probe Tool. We will replace it with a new one free of charge. 3. YOU ARE UNSURE. If you are uncomfortable doing this inspection yourself or have doubt whether the Probe Tool is good, we will send you a shipping label to return your existing Probe Tool. We will replace it with a new one free of charge. x-navtech.com | Customer Service: 1-267-436-0420 Ext. 2 | [email protected]
Quantity in commerce
102 units
Distribution
US Nationwide distribution and Foreign country of: Canada
Product codes and lots
UDI-DI: (01)00817421020995 (10) 2007010010 Lot number: 2007010010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0222-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0222-2023Class IIOngoingVoluntary: Firm initiated