LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral...
Integrity Implants Inc. began a recall of LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation... on . FDA classified it as Class II and gives the reason as "Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0418-2023
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integrity Implants, Inc.
- Product
- LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 50mm; o. 7.5mm x 55mm; p. 7.5mm x 60mm; q. 8.5mm x 35mm; r. 8.5mm x 40mm; s. 8.5mm x 45mm; t. 8.5mm x 50mm; u.8.5mm x 55mm; v. 8.5mm x 60mm; Used in spinal fixation surgery.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.
- Quantity in commerce
- 1,227 screws
- Distribution
- Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
- Product codes and lots
- a. 5.5mm x 35mm, Part Number: LS-N6SA5535, UDI: (01)00818613027716(10)02210117, lot Number: 2210117; Part Number: LS-N6SA7545, UDI: (01)00818613027884(10)02210130, Lot Number: 2210130. b. 5.5mm x 40mm, Part Number: LS-N6SA5540, UDI: (01)00818613027723(10)02210118, Lot Number: 2210118. c. 5.5mm x 45mm, Part Number: LS-N6SA5545, UDI: (01)00818613027730(10)02210119, Lot Number: 2210119. d. 5.5mm x 50mm, Part Number: LS-N6SA5550, UDI: (01)00818613027747(10)02210120, Lot Number: 2210120. e. 5.5mm x 55mm, Part Number: LS-N6SA5555, UDI: (01)00818613027754(10)02210121, Lot Number: 2210121. f. 6.5mm x 35mm, Part Number: LS-N6SA6535, UDI: (01)00818613027792(10)02210122, Lot Number: 2210122. g. 6.5mm x 40mm, Part Number: LS-N6SA6540, UDI: (01)00818613027808(10)02210123, Lot Number: 2210123. h. 6.5mm x 45mm, Part Number: LS-N6SA6545, UDI: (01)00818613027815(10)02210124, Lot Number: 2210124. i. 6.5mm x 50mm, Part Number: LS-N6SA6550, UDI: (01)00818613027822(10)02210125, Lot Number: 2210125. j. 6.5mm x 55mm, Part Number: LS-N6SA6555, UDI: (01)00818613027839(10)02210126, Lot Number: 2210126. k. 6.5mm x 60mm, Part Number: LS-N6SA6560, UDI: (01)00818613027846(10)02210127, Lot Number: 2210127. l. 7.5mm x 35mm, Part Number: LS-N6SA7535, UDI: (01)00818613027860(10)02210128, Lot Number: 2210128. m. 7.5mm x 40mm, Part Number: LS-N6SA7540, UDI: (01)00818613027877(10)02210129, Lot Number: 2210129. n. 7.5mm x 50mm, Part Number: LS-N6SA7550, UDI: (01)00818613027891(10)02210131, Lot Number: 2210131. o. 7.5mm x 55mm, Part Number: LS-N6SA7555, UDI: (01)00818613027907(10)02210132, Lot Number: 2210132. p. 7.5mm x 60mm, Part Number: LS-N6SA7560, UDI: (01)00818613027914(10)02210133, Lot Number: 2210133. q. 8.5mm x 35mm, Part Number: LS-N6SA8535, UDI: (01)00818613027938(10)02210134, Lot Number: 2210134. r. 8.5mm x 40mm, Part Number: LS-N6SA8540, UDI: (01)00818613027945(10)02210135, Lot Number: 2210135. s. 8.5mm x 45mm, Part Number: LS-N6SA8545, UDI: (01)00818613027952(10)02210136, Lot Number: 2210136. t. 8.5mm x 50mm, Part Number: LS-N6SA8550, UDI: (01)00818613027969(10)02210137, Lot Number: 2210137. u.8.5mm x 55mm, Part Number: LS-N6SA8555, UDI: (01)00818613027976(10)02210138, Lot Number: 2210138. v. 8.5mm x 60mm, Part Number: LS-N6SA8560, UDI: (01)00818613027983(10)03210142, Lot Number: 3210142.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0418-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0419-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0420-2023 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY