NexGen Option Stemmed Tibial Component Size 8, Item Number 00-5986-057-02

Zimmer GmbH / initiated 2022-12-06 / Open, Classified

Zimmer, Inc. began a recall of NexGen Option Stemmed Tibial Component Size 8, Item Number 00-5986-057-02 on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0471-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
NexGen Option Stemmed Tibial Component Size 8, Item Number 00-5986-057-02
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.
FDA root cause
Under Investigation by firm
Action
Urgent Medical Device Recall notification letters dated 12/6/22 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. It is recommended that for patients implanted with the NexGen Stemmed Option Tibial Component with either the LPS Flex or LPS Flex GSF Femoral Components, maintain an appropriate index of suspicion for patients with any new pain, inability to bear weight, swelling or instability of the knee. The UK NJR data suggests that tibial component loosening is a key cause of the increased TKR revisions. If your patient experiences any new pain or other symptoms related to the TKR, it is recommended that additional clinical and/or radiographic follow-up is completed. 3. If an impacted
Quantity in commerce
284
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
All Unexpired Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0471-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0466-2023Class IIOngoingVoluntary: Firm initiated
Z-0467-2023Class IIOngoingVoluntary: Firm initiated
Z-0468-2023Class IIOngoingVoluntary: Firm initiated
Z-0469-2023Class IIOngoingVoluntary: Firm initiated
Z-0470-2023Class IIOngoingVoluntary: Firm initiated
Z-0471-2023Class IIOngoingVoluntary: Firm initiated