Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Maquet Cardiovascular Us Sales, LLC / initiated 2022-11-07 / Open, Classified / root cause: design

Getinge Usa Sales Inc began a recall of Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900) on . FDA classified it as Class II and gives the reason as "The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0422-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.
Root cause tag
design
FDA root cause
Component design/selection
Action
On 11/7/22 the firm notified customers via Urgent Medical Device Correction letters. Customers were instructed to immediately identify whether they had affected Flow-c or Flow-e anesthesia systems on hand. Customers who have affected devices may continue to use the device ensuring that a system checkout procedure is performed per the Operators Manual. The firm will replace the Power backup battery free of charge for customers with affected anesthesia systems. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9GETUSA / (888) 943 8872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity in commerce
65 US
Distribution
US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.
Product codes and lots
Flow-c UDI-DI: 7325710010457 Lot/Serial Numbers 4003 4777 4783 5024 5025 5026 5027 5028 5043 5044 5045 5046 5047 5048 5049 5050 5078 Flow-e UDI-DI: 7325710009765 Lot/Serial Numbers 50012 50013 50014 50015 50016 50017 50030 50031 50033 50034 50035 50036 50037 50038 50039 50040 50041 50042 50043 50046 50047 50048 50051 50052 50053 50054 50055 50056 50057 50058 50059 50060 50061 50062 50069 50070 50071 50072 50073 50074 50084 50085 50086 50087 50088 50089 50090 50091 50092 50093 50094 50096 50097 50108 50109 50110 50111 50132

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0422-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0422-2023Class IIOngoingVoluntary: Firm initiated