EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base
MicroPort Orthopedics Inc. began a recall of EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base on . FDA classified it as Class II and gives the reason as "One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0275-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Microport Orthopedics Inc.
- Product
- EVOLUTION MP Tibial Bases, Size 2+ LEFT, REF ETPKN2PL, knee tibial base
- Product code
- HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer
- Reason for recall
- One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- MicroPort Orthopedics Disseminated Urgent Field Safety Notices to its consignees on October 28, 2022 via email. The notices explained the issue, the risk, and requested the following: "ACTIONS TO BE TAKEN BY THE USER: Our records indicate that you did receive the above referenced product: - Immediately check inventory and quarantine all subject products - COMPLETE AND RETURN the attached FSCA Acknowledgement - Inform MicroPort Orthopedics of any adverse event immediately - Return any affected product to MicroPort Orthopedics, please see your local distributor for details TRANSMISSION OF THIS NOTICE: This notice needs to be passed on to all those who need to be aware within your organization. CONTACT REFERENCE PERSON: For questions or additional information please contact: MicroPort Scientific Coperatief U.A. Phone: +31 20 545 01 00 Email: [email protected]"
- Quantity in commerce
- 24 units
- Distribution
- International Distribution to countries of: Germany, Italy, France, Belgium, Austria, Greece, Romania, Japan, and P.R. China.
- Product codes and lots
- UDI/DI 0 0192629 08287 8, Lot 1916559
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0275-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0276-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.