Z-0420-2023Class II

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle...

Integrity Implants, Inc. / initiated 2022-01-27 / Terminated / root cause: design

Integrity Implants Inc. began a recall of LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis... on . FDA classified it as Class II and gives the reason as "Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0420-2023
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integrity Implants, Inc.
Product
LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 6.5mm x 35mm; f. 6.5mm x 40mm; g. 6.5mm x 45mm; h. 6.5mm x 50mm; i. 6.5mm x 55mm; j. 7.5mm x 35mm; k. 7.5mm x 40mm; l. 7.5mm x 45mm; m. 7.5mm x 50mm; n. 7.5mm x 55mm; o. 8.5mm x 35mm; p. 8.5mm x 40mm; q. 8.5mm x 45mm; r. 8.5mm x 50mm; s.8.5mm x 55mm; t. 9.5mm x 35mm; u. 9.5mm x 40mm; v. 9.5mm x 45mm; w. 9.5mm x 50mm; x. 9.5mm x 55mm; Used in spinal fixation surgery.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
Root cause tag
design
FDA root cause
Device Design
Action
Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.
Quantity in commerce
1,369 screws
Distribution
Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
Product codes and lots
a. 5.5mm x 35mm, Part Number: LS-N6MG5535, UDI: (01)00818613028898(10)02210093, lot Number: 2210093; UDI: (01)00818613028898(10)10200053, Lot Number: 10200053. b. 5.5mm x 40mm, Part Number: LS-N6MG5540, UDI: (01)00818613028904(10)02210094, lot Number: 2210094; UDI: (01)00818613028904(10)10200054, Lot Number: 10200054. c. 5.5mm x 45mm, Part Number: LS-N6MG5545, UDI: (01)00818613028911(10)02210095, lot Number: 2210095; UDI: (01)00818613028911(10)10200055, Lot Number: 10200055. d. 5.5mm x 50mm, Part Number: LS-N6MG5550, UDI: (01)00818613028928(10)02210096, lot Number: 2210096; UDI: (01)00818613028928(10)10200056, Lot Number: 10200056. e. 6.5mm x 35mm, Part Number: LS-N6MG6535, UDI: (01)00818613028973(10)02210097, lot Number: 2210097; UDI: (01)00818613028973(10)10200057, Lot Number: 10200057; UDI: (01)00818613028973(10)10200057-1, Lot Number: 10200057-1. f. 6.5mm x 40mm, Part Number: LS-N6MG6540, UDI: (01)00818613028980(10)02210098, lot Number: 2210098; UDI: (01)00818613028980(10)10200058, Lot Number: 10200058; UDI: (01)00818613028980(10)10200058-1, Lot Number: 10200058-1. g. 6.5mm x 45mm, Part Number: LS-N6MG6545, UDI: (01)00818613028997(10)02210099, lot Number: 2210099; UDI: (01)00818613028997(10)10200059, Lot Number: 10200059; UDI: (01)00818613028997(10)10200059-1, Lot Number: 10200059-1. h. 6.5mm x 50mm, Part Number: LS-N6MG6550, UDI: (01)008181613029000(10)02210100, lot Number: 2210100; UDI: (01)00818613029000(10)10200060, Lot Number: 10200060; UDI: (01)00818613029000(10)10200060-1, Lot Number: 10200060-1. i. 6.5mm x 55mm, Part Number: LS-N6MG6555, UDI: (01)00818613029017(10)02210101, lot Number: 2210101; UDI: (01)00818613029017(10)10200061, Lot Number: 10200061. j. 7.5mm x 35mm, Part Number: LS-N6MG7535, UDI: (01)00818613026528(10)02210102, lot Number: 2210102; UDI: (01)00818613026528(10)10200062, Lot Number: 10200062. k. 7.5mm x 40mm, Part Number: LS-N6MG7540, UDI: (01)00818613026535(10)02210103, lot Number: 2210103; UDI: (01)00818613026535(10)10200063, Lot Number: 10200063; UDI: (01)00818613026535(10)06210309, Lot Number: 6210309. l. 7.5mm x 45mm, Part Number: LS-N6MG7545, UDI: (01)00818613026542(10)02210104, lot Number: 2210104; UDI: (01)00818613026542(10)10200064, Lot Number: 10200064; UDI: (01)00818613026542(10)10200064-1, Lot Number: 10200064-1. m. 7.5mm x 50mm, Part Number: LS-N6MG7550, UDI: (01)00818613026559(10)02210105, lot Number: 2210105; UDI: (01)00818613026559(10)10200065, Lot Number: 10200065; UDI: (01)00818613026559(10)10200065-1, Lot Number: 10200065-1. n. 7.5mm x 55mm, Part Number: LS-N6MG7555, UDI: (01)00818613026566(10)02210106, lot Number: 2210106; UDI: (01)00818613026566(10)10200066, Lot Number: 10200066. o. 8.5mm x 35mm, Part Number: LS-N6MG8535, UDI: (01)00818613026597(10)02210107, lot Number: 2210107; UDI: (01)00818613026597(10)10200067, Lot Number: 10200067. p. 8.5mm x 40mm, Part Number: LS-N6MG8540, UDI: (01)00818613026603(10)02210108, lot Number: 2210108; UDI: (01)00818613026603(10)10200068, Lot Number: 10200068. q. 8.5mm x 45mm, Part Number: LS-N6MG8545, UDI: (01)00818613026610(10)02210109, lot Number: 2210109; UDI: (01)00818613026610(10)10200069, Lot Number: 10200069. r. 8.5mm x 50mm, Part Number: LS-N6MG8550, UDI: (01)00818613026627(10)02210110, lot Number: 2210110; UDI: (01)00818613026627(10)10200070, Lot Number: 10200070. s. 8.5mm x 55mm, Part Number: LS-N6MG8555, UDI: (01)00818613026634(10)02210111, lot Number: 2210111; UDI: (01)00818613026634(10)10200071, Lot Number: 10200071. t. 9.5mm x 35mm, Part Number: LS-N6MG9535, UDI: (01)00818613026658(10)02210112, lot Number: 2210112; UDI: (01)00818613026658(10)10200072, Lot Number: 10200072. u. 9.5mm x 40mm, Part Number: LS-N6MG9540, UDI: (01)00818613026665(10)02210113, lot Number: 2210113; UDI: (01)00818613026665(10)10200073, Lot Number: 10200073. v. 9.5mm x 45mm, Part Number: LS-N6MG9545, UDI: (01)00818613026672(10)02210114, lot Number: 2210114; UDI: (01)00818613026672(10)10200074, Lot Number: 10200074. w. 9.5mm x 50mm, Part Number: LS-N6MG9550, UDI: (01)00818613026689(10)02210115, lot Number: 2210115; UDI: (01)00818613026689(10)10200075, Lot Number: 10200075. x. 9.5mm x 55mm, Part Number: LS-N6MG9555, UDI: (01)00818613026696(10)02210116, lot Number: 2210116; UDI: (01)00818613026696(10)10200076, Lot Number: 10200076.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0420-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0418-2023Class IITerminatedVoluntary: Firm initiated
Z-0419-2023Class IITerminatedVoluntary: Firm initiated
Z-0420-2023Class IITerminatedVoluntary: Firm initiated