Z-0721-2023Class II

Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms)...

GE Healthcare / initiated 2022-10-28 / Open, Classified / root cause: software

GE Medical Systems, LLC began a recall of Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms) and data from various imaging sources on . FDA classified it as Class II and gives the reason as "There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0721-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms) and data from various imaging sources.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
GE Healthcare notified consignees on about 10/28/2022 via letter titled, "URGENT MEDICAL DEVICE CORRECTION - UPDATE." Consignees were instructed that the device can continued being used by following these instructions: 1) When a study is not present on the Study List, contact a GE Healthcare Service Representative for assistance in recovering the study. 2) In situations where study interpretation is time sensitive, interpret the study at the scanning unit or plan for an alternate means of reading studies for emergent cases. GE Healthcare will contact consignees to arrange for the correction. Consignees were also instructed to complete and return the acknowledgement form.
Quantity in commerce
1 unit
Distribution
Nationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.
Product codes and lots
Installed Product ID: YP3150-Centricity Universal Viewer-01639018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0721-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0721-2023Class IIOngoingVoluntary: Firm initiated
Z-0722-2023Class IIOngoingVoluntary: Firm initiated