Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32...

Datascope Corp. / initiated 2022-12-19 / Open, Classified / root cause: design

Datascope Corp. began a recall of Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35... on . FDA classified it as Class I and gives the reason as "The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0885-2023
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-32, 0998-UC-0800-33, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-0800-36, 0998-UC-0800-45, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55, 0998-UC-0800-65.
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
Root cause tag
design
FDA root cause
Device Design
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/19/2022 was sent to customers. It contains clinical guidance to augment current clinical recommendations of patient management strategies should an IAB perforation occur. Maintain established patient and device management strategies regarding timing to IAB catheter removal. User Actions to be taken include to examine inventory immediately to determine if affected units are on hand, if a blood back event is suspected, remove the Cardiosave from patient use, and report the event to appropriate biomedical engineering staff for inspection prior to next patient use, and ensure that all Cardiosave Intra-Aortic Balloon Pump users at your facility are aware of this notice. Customers were also instructed to sign and return the MEDICAL DEVICE CORRECTION - RESPONSE FORM and notify customers if affected units were further distributed. The Cardiosave IABP units distributed going forward will be added to the scope and newly affected consignee(s) will be notified of the field correction. Letter dated 11/21/23 titled THIS IS AN UPDATE TO THE PREVIOUSLY INITIATED ACTION URGENT MEDICAL DEVICE CORRECTION was sent to customers. This notification is an update to the previously initiated action, dated December 19, 2022, regarding potential for blood to enter into the Cardiosave Hybrid and Rescue IABPs. The update is primarily to share new recommendations, such as a new labeling addendum and to consider the benefits and risks of using Cardiosave IABP on a patient during transport. Instructions were the same as the first letter. Updated letters will be sent to OUS consignees, beginning on 07/31/2025 alerting consignees of an upcoming component correction and software update to resolve the previous issues. Letters will be distributed as approval is received for each country. Once approval is received in the US, US consignees will be notified.
Quantity in commerce
11,906 Total
Distribution
Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLIC, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, TRINIDAD AND TOBAGO, Tunisia, Turkey, Turkmenistan, Ukraine, UNITED ARAB EMIRATES, United Kingdom, Vietnam, Yemen, and Zambia.
Product codes and lots
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-00-0800-33, 10607567109053; 0998-00-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-31, 10607567109053; 0998-UC-0800-32, 10607567111117; 0998-UC-0800-33, 10607567109008; 0998-UC-0800-34, 10607567111940; 0998-UC-0800-35, 10607567109107; 0998-UC-0800-36, 10607567114187; 0998-UC-0800-45, 10607567108421; 0998-UC-0800-52, 10607567108438; 0998-UC-0800-53, 10607567108391; 0998-UC-0800-55, 10607567108414; 0998-UC-0800-65, 10607567113432.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0885-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0885-2023Class IOngoingVoluntary: Firm initiated
Z-0886-2023Class IOngoingVoluntary: Firm initiated