Z-0462-2023Class II

Philips MR 7700 System, Model Number 782120

Philips North America LLC / initiated 2022-11-22 / Open, Classified

Philips North America Llc began a recall of Philips MR 7700 System, Model Number 782120 on . FDA classified it as Class II and gives the reason as "Gradient Coil may act as a heat source with a potential to produce smoke and/or fire". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0462-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Philips MR 7700 System, Model Number 782120
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
FDA root cause
Under Investigation by firm
Action
An URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users 1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your impacted MR system(s) until Philips has installed the SmokeDetector Interlock. " Post a Do not use notice on or near the impacted MR system(s) " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. " Note: Once your system has the SmokeDetector Interlock installed, please follow the advice in steps 2 through 5. 2. Philips has installed a SmokeDetector Interlock feature on the systems listed in Appendix B. Immediately confirm if your system has the SmokeDetector Interlock installed by confirming your serial number is listed in Appendix B. " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. 3. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: " Remove patient from the system according to the instructions for use because
Quantity in commerce
566 total
Distribution
US Nationwide.
Product codes and lots
Model 782120: UDI-DI: 00884838104112; Serial Numbers: 65005 65006 65007 65008

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0462-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0459-2023Class IIOngoingVoluntary: Firm initiated
Z-0460-2023Class IIOngoingVoluntary: Firm initiated
Z-0461-2023Class IIOngoingVoluntary: Firm initiated
Z-0462-2023Class IIOngoingVoluntary: Firm initiated
Z-0463-2023Class IIOngoingVoluntary: Firm initiated