Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
Leica Microsystems, Inc. began a recall of Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system on . FDA classified it as Class II and gives the reason as "There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0476-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
- Product code
- EPT Microscope, Surgical
- Reason for recall
- There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm issued an URGENT FIELD SAFETY NOTICE to its consignees by email on 11/21/2022. The notice explained the issue and stated that a Leica Microsystems representative would be contacting each facility to arrange an update to the the affected devices and advised the following: Consignees can continue to use the system pending the field update; Ensure that all users of the affected microscope are aware of the Safety notes as described in section 3 of the respective user manual; Like any other instrument in the operating theater, the system may fail and Leica Microsystem (Schweiz) AG recommends that a backup system be kept available during the operation; Check before the operation : Connect the power cable. + Switch on the microscope. UPDATE 09/12/2023: Recall was expanded, and customers were sent the notification on about 09/11/2023. Each consignee is requested to acknowledge receipt of the Field Safety Notice letter outlining the potential manufacturing defect, perform an initial inspection, indicate the results of the initial inspection, and provide contact information for further Leica Microsystems Representative verification of conformity/defect.
- Quantity in commerce
- 38 units
- Distribution
- US Nationwide
- Product codes and lots
- ***EXPANSION 09/08/2023 UPDATE a. Premium Part Number 10448950; Serial Numbers: 160222001, 280622001, 50422001, 50422002, 91221001, 180622001, 170622001, 61221001, 130922001, 120623001, 150623001, 81022001, 200623001, 51022001, 61022001. b. Standard Part Number 10448976; Serial Numbers: 120322001, 300122001, 40222002, 301221001, 40122001, 140522001, 151221002, 81221002, 100222002, 300122002, 110322001, 50322001, 90622001, 120422001, 90522001, 100522001, 250122001, 100222001, 171022001, 80922001, 70623001, 30922001, 91022001.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0475-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0476-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0477-2023 | Class II | Ongoing | Voluntary: Firm initiated |