Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0760-2023Class II

Revaclear 400 Dialyzer

Baxter Healthcare Corporation / initiated 2022-11-23 / Open, Classified

Baxter Healthcare Corporation began a recall of Revaclear 400 Dialyzer. Product Code: 114746L. Used to treat chronic and acute renal failure by hemodialysis on . FDA classified it as Class II and gives the reason as "Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0760-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Revaclear 400 Dialyzer. Product Code: 114746L. Used to treat chronic and acute renal failure by hemodialysis.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer
FDA root cause
Process change control
Action
Baxter Healthcare sent a notification via USPS first class mail on about 11/23/2022 and to home patients on about 11/29/2022. Consignees were instructed to locate and remove all affected product from your inventory, contact Baxter Healthcare Center for Service to arrange for return and credit, and acknowledge the receipt of the notification by completing a reply form on the customer portal. Note: this letter describes two issues. This recall addresses Internal Blood Leaks being caused by a twisted gasket in the dialyzer. The second issue is linked to reports of Polyurethane Displacement in the dialyzer, which is addressed in a separate recall.
Quantity in commerce
3,610,944 units
Distribution
Worldwide distribution. US Nationwide including Guam and St. Thomas USVI; Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, and Trinidad and Tobago.
Product codes and lots
UDI-DI: 07332414124076; Product Code: 114746L; Lot Numbers: C421222601, C421222701, C421222801, C421222901, C421223001, C421223101, C421223201, C421223301, C421223401, C421223501, C421223601, C421223701, C421223801, C421223901, C421224001, C421224101, C421224201, C421224301, C421224401, C421224601, C421224701, C421224801, C421224901, C421225001, C421225101, C421225201, C421225301, C421225401, C421225501, C421225601, C421225701, C421225801, C421225901, C421226001, C421226101, C421226201, C421226301, C421226401, C421226501, C421226601, C421226701, C421226801, C421226901, C421227001, C421227101, C421227201, C421227301, C421227401, C421227501, C421227601, C421227701, C421227801, C421227901, C421228001, C421228101, C421228201, C421228301, C421228401, C421228501, C421228601, C421228701, C421228801, C421228901, C421229001, C421229101, C421229201, C421229301, C421229401, C421231001, C421240001, C421240101, C421240201, C421240301, C421240401, C421240501, C421240601, C421240701, C421240801, C421240901, C421241001, C421241101, C421241201, C421241301, C421241401, C421241501, C421241601, C421241701, C421241801, C421241901, C421242001, C421242101, C421242201, C421242301, C421242401, C421242501, C421242601, C421242801, C422200101, C422200201, C422200301, C422200501, C422200601, C422200701, C422200801, C422200901, C422201001, C422201101, C422201201, C422201701, C422201801, C422202101, C422202201, C422202301, C422205001, C422205101, C422205201, C422205301, C422205701, C422205801, C422205901, C422206001, C422206101, C422206201, C422206301, C422206401, C422206501, C422206601, C422206701, C422206801, C422206901, C422207001, C422207201, C422207301, C422207401, C422207501, C422207601, C422207701, C422207801, C422207901, C422208001, C422208101, C422208201, C422208301, C422208401, C422208501, C422208601, C422208701, C422208801, C422208901, C422209001, C422209101, C422209201, C422209301, C422209401, C422209501, C422209801, C422209901, C422210001, C422210101, C422210201, C422210301, C422210401, C422210501, C422210601, C422210701, C422211201, C422211301, C422211401, C422211501, C422211601, C422211701, C422211801, C422211901, C422212001, C422212101, C422212201, C422212301, C422212401, C422212501, C422212601

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0760-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0760-2023Class IIOngoingVoluntary: Firm initiated
Z-0761-2023Class IIOngoingVoluntary: Firm initiated