Z-0719-2023Class II
Oncentra Brachy radiation therapy planning system software
Elekta Limited / initiated 2022-12-05 / Open, Classified / root cause: software
Elekta, Inc. began a recall of Oncentra Brachy radiation therapy planning system software on . FDA classified it as Class II and gives the reason as "A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0719-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Elekta Limited
- Product
- Oncentra Brachy radiation therapy planning system software.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta issued an Important Field Safety Notice 806-01-BTP-003 dated November 2022 to all affected customers on 12/5/2022 via email. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The notice is to be posted in a place accessible to all users until this action is closed. Appropriate personnel working with the product are to be advised of the content of the letter. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no longer than 30 days from receipt. Update: The manufacturer has now released the solution to this issue for affected Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00. All affected customers will receive a software update in order to solve the issue. The solution was released for installation in the field on 20-MAR-23. Customers are being notified by email on 03/20/2023. Customers are being informed that the fix is available for download/request depending on their standard processes. Elekta service teams will pro-actively follow up with customers to insure the fix has been installed before the target completion date.
- Quantity in commerce
- 336 systems
- Distribution
- Distribution US nationwide, including Puerto Rico. There was no military/government distribution.
- Product codes and lots
- Version 4.0 and above with Applicator Modeling or Implant Modeling; UDI GTIN numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0719-2023 | Class II | Ongoing | Voluntary: Firm initiated |