Oncentra Brachy radiation therapy planning system software

Elekta Limited / initiated 2022-12-05 / Open, Classified / root cause: software

Elekta, Inc. began a recall of Oncentra Brachy radiation therapy planning system software on . FDA classified it as Class II and gives the reason as "A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0719-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Elekta Limited
Product
Oncentra Brachy radiation therapy planning system software.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
Root cause tag
software
FDA root cause
Software design
Action
Elekta issued an Important Field Safety Notice 806-01-BTP-003 dated November 2022 to all affected customers on 12/5/2022 via email. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The notice is to be posted in a place accessible to all users until this action is closed. Appropriate personnel working with the product are to be advised of the content of the letter. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no longer than 30 days from receipt. Update: The manufacturer has now released the solution to this issue for affected Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00. All affected customers will receive a software update in order to solve the issue. The solution was released for installation in the field on 20-MAR-23. Customers are being notified by email on 03/20/2023. Customers are being informed that the fix is available for download/request depending on their standard processes. Elekta service teams will pro-actively follow up with customers to insure the fix has been installed before the target completion date.
Quantity in commerce
336 systems
Distribution
Distribution US nationwide, including Puerto Rico. There was no military/government distribution.
Product codes and lots
Version 4.0 and above with Applicator Modeling or Implant Modeling; UDI GTIN numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0719-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0719-2023Class IIOngoingVoluntary: Firm initiated