Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical...
Randox Laboratories Ltd. began a recall of Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2682... on . FDA classified it as Class II and gives the reason as "Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0903-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Randox Laboratories, Limited
- Product
- Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2682 Control Level 1
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
- FDA root cause
- Under Investigation by firm
- Action
- Randox Laboratories Ltd issued via email to the US Distributor Randox Laboratories-US, Ltd., Kearneysville, West Virginia on 11/22/22. The US Distributor contacted the customers by letter on 11/28/22. The Urgent Medical Device Correction letter states reason for recall, health risk and action to take: Discard previous IFUs and download the updated IFUs from randox.com Review results generated with the affected batches in line with the clinical profile of the patient. Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
- Quantity in commerce
- 146 units US
- Distribution
- US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.
- Product codes and lots
- GTIN: 05055273204896 Lot Number/Exp. Date: 584LPC 28th Jun 23; 600LPC 28th Nov 23; 611LPC 28th Sep 24;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0903-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0904-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0905-2023 | Class II | Ongoing | Voluntary: Firm initiated |