Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0910-2023Class II

Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT...

Medrad, Inc. / initiated 2022-11-30 / Open, Classified / root cause: packaging

Bayer Medical Care, Inc. began a recall of Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material number: 86566621 Intended to... on . FDA classified it as Class II and gives the reason as "Black specks may be present on the packaged needles". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0910-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medrad, Inc.
Product
Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material number: 86566621 Intended to deliver accurate doses of 18F Fluorodeoxyglucose (FDG) or 18F Sodium Fluoride (NaF) radiopharmaceuticals and commonly used flushing solutions to patients during molecular imaging (PET Positron Emission Tomography - nuclear medicine) diagnostic procedures.
Product code
FRN Pump, Infusion
Reason for recall
Black specks may be present on the packaged needles
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
Bayer US issued Urgent Medical Device Recall (RECALL ID: SA-22-RAD-05 ) Letter on 11/30/22 via email. Letter states reason for recall, health risk and action to take: 1. Review your current inventory for affected Lot 60272154 and quarantine, as appropriate. 2. Complete the included response form and submit through email to [email protected] or contact our Bayer Customer Service team at 1-800-633-7231, Press 0, then option 3. Please complete the Customer Response Form regardless of whether you have affected product or not. This action will assist us in tracking all impacted product. 3. You will receive a Return Goods Authorization (RGA) number to return any affected kits remaining in your inventory for a credit. 4. Place the affected product in a box and label the outside of the container with the RGA number in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return affected product as soon as possible 5. Upon Bayer receipt of returned product, you will receive a credit on your account. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Quantity in commerce
2840 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, IN, MN, NY, OH, TX, VA and the country of China.
Product codes and lots
UDI: (01)00616258021478(11)201209(17)231209(10)60272154 Batch number: 60272154 Expiration date: December 9, 2023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0910-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0910-2023Class IIOngoingVoluntary: Firm initiated