Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457
Integra LifeSciences Corp. began a recall of Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457 on . FDA classified it as Class II and gives the reason as "Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0909-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457
- Product code
- HBA Neurosurgical Paddie
- Reason for recall
- Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On December 2, 2022, the firm notified customers via "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letters. Customers were instructed to immediately remove units of affected product from service. Customers should complete the Acknowledgement Form and return to the recalling company. Once received, Customer Service will contact the customer to aid them in returning the affected product. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]
- Quantity in commerce
- 168 US; 3 OUS
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, IN, KY, MA, MI, MO, NC, NJ, NY, OH, PA, TX, VA, WV and the countries of Chile, Australia, New Zealand, France, Romania, United Kingdom, and Ireland.
- Product codes and lots
- UDI-DI 10381780514961 Lot Number: 6396457 Expiration date 06/30/2027