Z-1103-2023Class II

SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

OLYMPUS Corporation of the Americas / initiated 2022-12-13 / Open, Classified

Olympus Corporation of the Americas began a recall of SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D) on . FDA classified it as Class II and gives the reason as "A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1103-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to troubleshoot the device
FDA root cause
Under Investigation by firm
Action
Olympus issued URGENT - Medical Device Field Corrective Action letter on 12/13/22. Letter states reason for recall, health risk and action to take: 1. Notify your users about this issue and stop using the single-use bronchoscopes (models INS-7100 or INS-7130) with SYS-4000 system until cable is replaced. Your users can continue to use the SYS-4000 using a reusable bronchoscope, with its respective video processor, without affecting their electromagnetic navigation (EMN) procedural experience. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter the recall number 0418 Provide your contact information as indicated in the portal. An Olympus Field Service Representative will contact you to schedule a site visit to replace the DVI cable. 3. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 to report complaints. Contact (647) 999-3203 or at [email protected] for any additional information or support.
Quantity in commerce
13 units
Distribution
FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA
Product codes and lots
UDI: 00815686020101 Serial Numbers: 85330200430, 87896220519, 87899220524, 87379210921, 87835220505, 87832220502, 87767220420, 87777220415, 87827220427, 87736220412, 87891220517, 87864220509, 87738220412

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1103-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1103-2023Class IIOngoingVoluntary: Firm initiated