Z-0952-2023Class II

cobas infinity central lab, Material Number 07154003001

Roche Diagnostics GmbH / initiated 2022-12-12 / Open, Classified / root cause: software

Roche Diagnostics Operations, Inc. began a recall of cobas infinity central lab, Material Number 07154003001 on . FDA classified it as Class II and gives the reason as "A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0952-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Roche Diagnostics GmbH
Product
cobas infinity central lab, Material Number 07154003001
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions.
Root cause tag
software
FDA root cause
Software design
Action
On December 12, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were informed of the product issue. A Roche Diagnostics representative will contact you in the near future to assess your cobas infinity configuration and to determine if it meets all pre-conditions. If so, your Roche representative will work with you to implement a workaround immediately in addition to working with you to apply a more permanent solution via a software patch once available. Customers should consult with a physician or pathologist at your facility to determine clinical implications specific to your patients. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about this recall.
Quantity in commerce
224 US; 2,797 worldwide
Distribution
Domestic distribution nationwide. Foreign distribution pending.
Product codes and lots
cobas infinity central lab / cobas infinity core license using software versions (2.5.x , 3.01.x, 3.02.x, and 3.03.x) that have been configured with all of the following pre-conditions will be impacted: 1. Having Order ID configuration to include the date (Not Common in U.S. Configurations) 2. Having a host configured by Order rather than Tube (Not Common in U.S..Configurations) 3. Having message configured to receive registration date of the order using ExtRegisterDate and ExtRegisterHour. 4. Having HCA configuration enabled for " Reject orders if not possible to assign External ID" and the radio button option set to "To at least one test" UDI-DI 04015630936007

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0952-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0952-2023Class IIOngoingVoluntary: Firm initiated