Z-1037-2023Class II

Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized...

TURBETT SURGICAL, Inc / initiated 2022-12-05 / Open, Classified / root cause: design

Turbett Surgical, Inc. began a recall of Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider on . FDA classified it as Class II and gives the reason as "Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1037-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
TURBETT SURGICAL, Inc
Product
Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider
Product code
FRG Wrap, Sterilization
Reason for recall
Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
Root cause tag
design
FDA root cause
Device Design
Action
***UPDATE*** Turbett Surgical issued via E-mails Product Alert directly to the Hospital departments utilizing the device on 12-27-2022. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Immediately discontinue the use of the specific devices exhibiting one or more failure(s) listed below. " Missing side supports " Broken or cracked side tab support " Backwall weld separation " Side weld separation " Crack in the backwall seam 2. Email ([email protected]) with the following information. " Hospital name " Serial number(s) " Type of failure that your container exhibited 3. Turbett Surgical will contact you to organize pick up of the damaged unit and shipment of a replacement unit. If your container does not exhibit any of these failures, please continue to monitor and inspect the condition of the container and follow the Instructions For Use (IFU) as you continue utilizing your container. 4. If your container does not exhibit any of these failures, please email [email protected] stating that your container(s) do not exhibit any of these failures. Please continue to monitor and inspect your devices prior to assembling for sterilization AS WELL AS after sterilization and cooling to ensure expected outcomes. Reminder to follow the Instructions For Use (IFU) as you continue utilizing your TS-1200 container. If you have any questions about the content of this letter, please contact Mark Petras at [email protected]. ________________________________________________________________________ Turbett Surgical issued Emails to the Hospital department on 12/7/2022. Sales representatives were notified on 12/5/22 to switch out the other container. Letter states reason for recall, health risk and action to take: 1. Immediately verify that the specific serial numbered devices listed below, continues to operate without creating wet loads or the sterilization process is not compromised.
Quantity in commerce
61 units
Distribution
US Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.
Product codes and lots
GTIN: 00859464007281 Lot Number: All Lot Numbers Serial Number: All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1037-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1037-2023Class IIOngoingVoluntary: Firm initiated
Z-1038-2023Class IIOngoingVoluntary: Firm initiated