Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0773-2023Class I

RSDL (Reactive Skin Decontamination Lotion) Kit, NSN 6505-21-912-5229. Used to remove and/or neutralize...

Serb Medical Solutions Inc. / initiated 2022-11-24 / Terminated / root cause: packaging

Emergent Protective Products USA Inc began a recall of RSDL (Reactive Skin Decontamination Lotion) Kit, NSN 6505-21-912-5229. Used to remove and/or neutralize chemical warfare agents and T-2 Toxin from the skin on . FDA classified it as Class I and gives the reason as "Complaints of leaking packets were received". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0773-2023
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Serb Medical Solutions Inc.
Product
RSDL (Reactive Skin Decontamination Lotion) Kit, NSN 6505-21-912-5229. Used to remove and/or neutralize chemical warfare agents and T-2 Toxin from the skin.
Product code
MAC Decontamination Kit
Reason for recall
Complaints of leaking packets were received.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Emergent sent a RECALL NOTIFICAITON LETTER to its consignees by email on 11/24/2022. The notice explained the issue and requested that the consignee immediately hold the product and notify the firm if any product was forwarded to another organization for further distribution. The email response form provided self inspection instructions and to segregate failed packets and store them in a quarantine location to prevent distribution. Further information and materials were provided to return defective product to Emergent. Packets that have passed inspection criteria can be placed back into active inventory A follow up email recall communication was issued on 12/02/2022 as follows: "Reason for this Additional Notification: To provide information for product replacement and return. To provide information on Sub-Recall procedures. There is potential that packets that have passed visual inspection may have a seal that could be compromised under the certain conditions. These packets will be replaced with new product. In the interim period prior to receipt of new product, the packets that have passed your visual inspection can remain in your hold inventory. Instructions: Please follow the following steps: Complete actions as per the initial notification letter provided on 24 NOV 2022. After the completion of the visual inspection, packets that have passed should be placed into hold inventory. Receive replacement product. As soon as available, Emergent will ship new product to replace the entire initial shipment of the batch associated with the recall. Once the new inventory is received, all remaining packets from the recalled batch (Passed and Failed) should be returned to Emergent. Initiate Sub-Recalls to all customers you have distributed product to. Should you have any questions regarding this product recall, please contact: Brent Sadler, Sr. Director, Quality, [email protected], 204-396-3815"
Quantity in commerce
4000 units
Distribution
Distribution to US state of VA (distributor), Canada, and Switzerland
Product codes and lots
UDI-DI: 00817129020013, Lot numbers: 23005060, 23005061, 23005062. (01)00817129020013 (10) 23005060 (17) 270718 (F5408ENG)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0773-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0773-2023Class ITerminatedVoluntary: Firm initiated