Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part...
Getinge Usa Sales Inc began a recall of Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400 on . FDA classified it as Class I and gives the reason as "Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0959-2023
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- On December 22, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to examine their inventory immediately to determine if they have any of the affected serial numbers. If you have an affected system running on software version 4.8.0 through 4.8.3, you may continue use of the device until the software is updated with the following considerations: 1. If during use, the device is switched from Automatic to Manual and it is found that the manual bag is empty the user can activate the Emergency ventilation or turn off/on the device to reset the disabled valves function. 2. Users may also use the resuscitator that is available in the Operating Room. Getinge will update the software to correct the issue free of charge. You can arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9GETUSA / (888) 943 8872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Quantity in commerce
- 8,016 Total (7,766 OUS; 250 US)
- Distribution
- US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.
- Product codes and lots
- Part # UDI 6677200 7325710001349 Serial Numbers 1202-8027 Part # UDI 6677300 7325710001363 Serial Numbers 1058-1201 Part # UDI 6677400 7325710001318 Serial Numbers 2034-8446 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0959-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0960-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0961-2023 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
