Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1207-2023Class II

BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis...

CareFusion 303, Inc / initiated 2023-02-14 / Open, Classified / root cause: software

CareFusion 303, Inc. began a recall of BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all... on . FDA classified it as Class II and gives the reason as "Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1207-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all with BD Pyxis ES version 1.6.1 drawer firmware (version 1.10.2.5). Pyxis MedStation Enterprise Server (ES) System is an automated medication dispensing system supporting decentralized medication management in a pharmaceutical environment.
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.
Root cause tag
software
FDA root cause
Software change control
Action
On 01/14/23, correction notices were mailed to directors of pharmacy informing them to take the following actions: 1) Ensure this letter and contents are forwarded and/or discussed with all appropriate personnel to ensure proper awareness of the issue. 2) If you are experiencing any issues, such as drawer failures, call BD technical support and refer to user guides for additional information and troubleshooting procedures. a. BD Pyxis MedStation ES user guide, refer to sections "Recover storage space" (Chapter 9, pg. 76) and "Emergency medication removal" (Appendix 11, pg. 119). See references 1 and 2 below for additional details. b. BD Pyxis PAS ES user guide, refer to sections "Emergency access procedures" (Appendix A, pg. 77) and "Storage spaces are not responsive" (Appendix B, pg. 81). See references 3 and 4 below for additional details. 3) Complete and return the Customer Response Form. The firm will take the following actions: 1) All customers who may have affected devices will be notified. 2) Upon receipt of completed response forms, the firm will coordinate the firmware upgrade with customers. Technical Support: 1-800-727-6102, Phone hours: 24 hours, 7 days a week
Quantity in commerce
7914
Distribution
Worldwide - US Nationwide distribution including in the states of NY, PA, OH, TX, FL, KY, IL, DE, AK, WI, KS, LA, WA, AZ, CO, MI, IN, TN, SD, AR, MO, CA, NV, MA, SC, VA, WV, CT, ME, NE, RI, GA, NC, IA, MN, AL, OK, MS, NM, NH, VT, WY, OR, ID, NJ, UT, DC, HI, MD, MT, ND and the country of Canada.
Product codes and lots
REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1207-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1207-2023Class IIOngoingVoluntary: Firm initiated