BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis...
CareFusion 303, Inc. began a recall of BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all... on . FDA classified it as Class II and gives the reason as "Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1207-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CareFusion 303, Inc
- Product
- BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES Tower (Med ES Tower), REF: 352; all with BD Pyxis ES version 1.6.1 drawer firmware (version 1.10.2.5). Pyxis MedStation Enterprise Server (ES) System is an automated medication dispensing system supporting decentralized medication management in a pharmaceutical environment.
- Product code
- BRY Cabinet, Table And Tray, Anesthesia
- Reason for recall
- Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 01/14/23, correction notices were mailed to directors of pharmacy informing them to take the following actions: 1) Ensure this letter and contents are forwarded and/or discussed with all appropriate personnel to ensure proper awareness of the issue. 2) If you are experiencing any issues, such as drawer failures, call BD technical support and refer to user guides for additional information and troubleshooting procedures. a. BD Pyxis MedStation ES user guide, refer to sections "Recover storage space" (Chapter 9, pg. 76) and "Emergency medication removal" (Appendix 11, pg. 119). See references 1 and 2 below for additional details. b. BD Pyxis PAS ES user guide, refer to sections "Emergency access procedures" (Appendix A, pg. 77) and "Storage spaces are not responsive" (Appendix B, pg. 81). See references 3 and 4 below for additional details. 3) Complete and return the Customer Response Form. The firm will take the following actions: 1) All customers who may have affected devices will be notified. 2) Upon receipt of completed response forms, the firm will coordinate the firmware upgrade with customers. Technical Support: 1-800-727-6102, Phone hours: 24 hours, 7 days a week
- Quantity in commerce
- 7914
- Distribution
- Worldwide - US Nationwide distribution including in the states of NY, PA, OH, TX, FL, KY, IL, DE, AK, WI, KS, LA, WA, AZ, CO, MI, IN, TN, SD, AR, MO, CA, NV, MA, SC, VA, WV, CT, ME, NE, RI, GA, NC, IA, MN, AL, OK, MS, NM, NH, VT, WY, OR, ID, NJ, UT, DC, HI, MD, MT, ND and the country of Canada.
- Product codes and lots
- REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1207-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.