Z-1016-2023Class II
Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro...
Future Diagnostics Solutions B.V. / initiated 2022-12-06 / Open, Classified
Future Diagnostics Solutions B.V. began a recall of Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1016-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Future Diagnostics Solutions B.V.
- Product
- Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.
- Product code
- CEW Radioimmunoassay, Parathyroid Hormone
- Reason for recall
- Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.
- FDA root cause
- Under Investigation by firm
- Action
- On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to [email protected].
- Quantity in commerce
- 342 kits
- Distribution
- Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom
- Product codes and lots
- Catalog Number: 4K-IPT-00 UDI-DI: 08719326045214-230915 Lot Numbers: M78559 M78560 M78636
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1016-2023 | Class II | Ongoing | Voluntary: Firm initiated |