Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0932-2023Class II

PHS Lacertaion Tray, Sterile, 20 per case, Manufactured for Professional Hosptial Supply, Inc

Gyn-Disposables, Inc. / initiated 2022-12-15 / Open, Classified / root cause: manufacturing process

Stradis Healthcare began a recall of PHS Lacertaion Tray, Sterile, 20 per case, Manufactured for Professional Hosptial Supply, Inc on . FDA classified it as Class II and gives the reason as "Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0932-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Gyn-Disposables, Inc.
Product
PHS Lacertaion Tray, Sterile, 20 per case, Manufactured for Professional Hosptial Supply, Inc.
Product code
PKD Laceration Tray
Reason for recall
Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Quantity in commerce
300 units
Distribution
US Nationwide distribution in the states of FL, IL, IN, MA, MS, NV, OH, TN, TX, WA.
Product codes and lots
Assembly Number 477944, UDI-DI: M7524779441, Lot Numbers: 220974270, 220915145

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0932-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0931-2023Class IIOngoingVoluntary: Firm initiated
Z-0932-2023Class IIOngoingVoluntary: Firm initiated
Z-0933-2023Class IIOngoingVoluntary: Firm initiated
Z-0934-2023Class IIOngoingVoluntary: Firm initiated
Z-0935-2023Class IIOngoingVoluntary: Firm initiated
Z-0936-2023Class IIOngoingVoluntary: Firm initiated
Z-0937-2023Class IIOngoingVoluntary: Firm initiated
Z-0938-2023Class IIOngoingVoluntary: Firm initiated
Z-0939-2023Class IIOngoingVoluntary: Firm initiated
Z-0940-2023Class IIOngoingVoluntary: Firm initiated
Z-0941-2023Class IIOngoingVoluntary: Firm initiated
Z-0942-2023Class IIOngoingVoluntary: Firm initiated
Z-0943-2023Class IIOngoingVoluntary: Firm initiated
Z-0944-2023Class IIOngoingVoluntary: Firm initiated
Z-0945-2023Class IIOngoingVoluntary: Firm initiated
Z-0946-2023Class IIOngoingVoluntary: Firm initiated
Z-0947-2023Class IIOngoingVoluntary: Firm initiated
Z-0948-2023Class IIOngoingVoluntary: Firm initiated