Tacy Medical, Inc Fluid Filter Transfer Kit, Sterile, 60 kits per bag, 5 bags per case. Manufactured for...
Stradis Healthcare began a recall of Tacy Medical, Inc Fluid Filter Transfer Kit, Sterile, 60 kits per bag, 5 bags per case. Manufactured for Tracy Medical, Inc on . FDA classified it as Class II and gives the reason as "Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0945-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Gyn-Disposables, Inc.
- Product
- Tacy Medical, Inc Fluid Filter Transfer Kit, Sterile, 60 kits per bag, 5 bags per case. Manufactured for Tracy Medical, Inc.
- Product code
- LRS I.V. Start Kit
- Reason for recall
- Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
- Quantity in commerce
- 1,175 units
- Distribution
- US Nationwide distribution in the states of FL, IL, IN, MA, MS, NV, OH, TN, TX, WA.
- Product codes and lots
- Assembly Number TM.FFTK, UDI-DI: M752TMFFTK0, Lot Numbers: 220954416, 222418904, 221056552, 220070456, 220400116, 220411513, 220952568, 220943346, 221727862, 222709611, 222701385
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0931-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0932-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0933-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0934-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0935-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0936-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0937-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0938-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0939-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0940-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0941-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0942-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0943-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0944-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0945-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0946-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0947-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0948-2023 | Class II | Ongoing | Voluntary: Firm initiated |