AxiEM" Non-Invasive Patient Tracker
Medtronic Navigation, Inc. began a recall of AxiEM" Non-Invasive Patient Tracker on . FDA classified it as Class II and gives the reason as "Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1039-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Navigation, Inc
- Product
- AxiEM" Non-Invasive Patient Tracker
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 12/15/2022, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via mail courier informing customers is recalling two specific lots of AxiEM Non-Invasive Patient Trackers due to the potential for verification failure of navigation instruments during a Functional Endoscopic Sinus Surgery procedure. Customer are instructed: 1. Do NOT use any impacted product. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below. 2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. Their Medtronic Sales Representative can assist in initiating the return of the product as necessary. 3. Complete the Customer Confirmation Form enclosed with this letter (even if they have no product to return), acknowledging that they have received this information, and email the completed form to Medtronic at [email protected]. 4. If the affected devices have already been utilized and/or discarded, the firm is still asking that customers complete and return the Customer Confirmation Form detailing that information. For questions, contact Medtronic Sales Representative or Medtronic Technical Services at 1-888-826-5603 or via email at [email protected].
- Quantity in commerce
- 1,867 devies
- Distribution
- Worldwide - U.S. distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The country of Japan.
- Product codes and lots
- Product Number/CFN: 9734887XOM UDI-Device Identifier (GTIN/UPN): 00763000588380 Lot Number: 220728I 220729 Product Number/CFN: 9734887XOM UDI-Device Identifier (GTIN/UPN): 00643169608252 Lot Number: 220728I
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1039-2023 | Class II | Ongoing | Voluntary: Firm initiated |