Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2...
Meridian Bioscience Inc began a recall of Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700 on . FDA classified it as Class II and gives the reason as "Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1008-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Meridian Bioscience Inc
- Product
- Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700
- Product code
- QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
- Reason for recall
- Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Meridian Bioscience issued Urgent Medical Device Recall on 12/15/22 via email. The letter states reason for recall, health risk and action to take: REQUIRED ACTIONS: " Inspect the rotor, ring, and chamber walls of the Revogene instrument for signs of leakage (e.g., dried spots). Contact Meridian Bioscience Technical Service if any signs of leakage were detected. " If leakage is suspected, clean the Revogene, MOCK PIEs, and surrounding workspace using the instructions in the Revogene operator's manual. " Review current inventory and segregate any remaining stock. " Discontinue use of any remaining stock. " Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return any remaining inventory to Meridian Bioscience Inc. for credit. " Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Complete this form even if you have no remaining inventory. Questions, please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at [email protected].
- Quantity in commerce
- 92 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, IN, KS, KY, LA, MI, MN, NC, NY, OR, TX.
- Product codes and lots
- UDI: N/A Lot Numbers 410700P019, 410700P020, 410700P021, 410700P022, 410700P023, 410700P024, 410700P027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1008-2023 | Class II | Ongoing | Voluntary: Firm initiated |