Z-1008-2023Class II

Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2...

Meridian Bioscience Inc / initiated 2022-12-15 / Open, Classified / root cause: manufacturing process

Meridian Bioscience Inc began a recall of Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700 on . FDA classified it as Class II and gives the reason as "Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1008-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Meridian Bioscience Inc
Product
Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700
Product code
QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
Reason for recall
Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Meridian Bioscience issued Urgent Medical Device Recall on 12/15/22 via email. The letter states reason for recall, health risk and action to take: REQUIRED ACTIONS: " Inspect the rotor, ring, and chamber walls of the Revogene instrument for signs of leakage (e.g., dried spots). Contact Meridian Bioscience Technical Service if any signs of leakage were detected. " If leakage is suspected, clean the Revogene, MOCK PIEs, and surrounding workspace using the instructions in the Revogene operator's manual. " Review current inventory and segregate any remaining stock. " Discontinue use of any remaining stock. " Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return any remaining inventory to Meridian Bioscience Inc. for credit. " Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Complete this form even if you have no remaining inventory. Questions, please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at [email protected].
Quantity in commerce
92 kits
Distribution
US Nationwide distribution in the states of AZ, CA, IN, KS, KY, LA, MI, MN, NC, NY, OR, TX.
Product codes and lots
UDI: N/A Lot Numbers 410700P019, 410700P020, 410700P021, 410700P022, 410700P023, 410700P024, 410700P027

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1008-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1008-2023Class IIOngoingVoluntary: Firm initiated