Z-0956-2023Class II

ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor

Siemens Medical Solutions USA, Inc. / initiated 2022-12-16 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor on . FDA classified it as Class II and gives the reason as "Mechanical connection between the tabletop and table base may be lost". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0956-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Mechanical connection between the tabletop and table base may be lost.
Root cause tag
design
FDA root cause
Device Design
Action
A customer letter dated 12/16/2022 was issued to affected consignees. Consignees are advised the following: -Before treatment is started, check if both transversal guide rails and the roller bearing at the end of each rail are in the correct position. -For the affected OR tables, an inspection including installation of a safety measure will be carried out. Siemens is currently developing this mitigation. Once correction is available, the service department will contact the consignees to arrange a date to perform the corrective action. Consignees are requested to acknowledge the recall letter and return the acknowledgement form to Siemens promptly.
Quantity in commerce
550 units
Distribution
Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Italy, Japan, Kazakhstan, Kuwait, Lebanon, Monaco, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, and Vietnam.
Product codes and lots
1) ARTIS pheno; Material #10849000; UDI/DI: 4056869046877 2) ARTIS icono biplane; Material #11327600; UDI/DI: 4056869063317 3) ARTIS icono floor; Material #11327700; UDI/DI: 4056869149325

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0956-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0956-2023Class IIOngoingVoluntary: Firm initiated