Z-0996-2023Class II
Vasoshield Syringe Packs, Model VH-5001
Maquet Cardiovascular LLC / initiated 2022-12-22 / Open, Classified
Maquet Cardiovascular, LLC began a recall of Vasoshield Syringe Packs, Model VH-5001 on . FDA classified it as Class II and gives the reason as "Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0996-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Maquet Cardiovascular LLC
- Product
- Vasoshield Syringe Packs, Model VH-5001
- Product code
- GBX Catheter, Irrigation
- Reason for recall
- Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.
- FDA root cause
- Under Investigation by firm
- Action
- URGENT MEDICAL DEVICE Removal notification letters dated 12/22/22 were sent to customers. Actions to be taken by the Customer: Our records indicate that you have received the Hemopro 2 with Vasoshield and Vasoshield Syringe from lots affected by this recall. Please, note that Distributed Affected Lot Numbers appear on the device shelf boxes only. Please examine your inventory immediately to determine if you have any of Hemopro 2 with Vasoshield and/or Vasoshield Syringe with the product code/lot numbers listed in this notice. LOT numbers can be found on the product label in the area circled in red. 1. Should you have any affected product lots as listed in this notification, please stop using and remove the complete device from areas of use. Distributed Affected Lot Number can be found on the device package. 2. Should you have un-used affected VH-4001 kits and/or VH-5001 (partial or full packs) you are eligible for credit. 3. Please forward this information to all current and potential Hemopro 2 with Vasoshield and Vasoshield Syringe users within your hospital / facility. 4. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. 5. Please contact Getinge Customer Service at 888-880-2874 between the hours of 6:00 a.m. and 5:00 p.m. Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. 6. Whether you have affected product or not, please complete and sign the attached MEDICAL DEVICE Removal - RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet Cardiovascular, LLC /Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (877)-804-2316. We apologize for any inconvenience this recall may cause. If you have any questions, please contact your Maquet Cardiovascular, LLC /Getinge representative or call the Geting
- Quantity in commerce
- 530 units US, 215 units OUS
- Distribution
- US Nationwide. Japan, Germany.
- Product codes and lots
- UDI-DI: 00607567700468; Serial Numbers: 25147126, 25151632, 25152389, 25152964, 25163138, 25162077
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0995-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0996-2023 | Class II | Ongoing | Voluntary: Firm initiated |