Z-1000-2023Class II

AXIOM Luminos TF, AXIOM Iconos R100, AXIOM Iconos R200 and AXIOM Luminos dRF

Siemens Medical Solutions USA, Inc. / initiated 2022-12-09 / Open, Classified / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of AXIOM Luminos TF, AXIOM Iconos R100, AXIOM Iconos R200 and AXIOM Luminos dRF on . FDA classified it as Class II and gives the reason as "There is potential for the footrest to detach from the patient table during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1000-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Luminos TF, AXIOM Iconos R100, AXIOM Iconos R200 and AXIOM Luminos dRF
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
There is potential for the footrest to detach from the patient table during use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A Customer Safety Advisory Notification (CSAN) was distributed to affected consignees beginning December 9, 2022. The CSAN was distributed to inform the customers about a preventative measure to exclude any potential risk of patient injury caused by the footrest detaching from the patient table during examination. Together with the CSAN, an addendum to the operator manual is distributed containing additional details and illustrations on how to securely attach the footrest to further improve the understanding of handling the footrest. Refer to the CSAN with addendum and copies of the Instructions for Use for each reported system.
Quantity in commerce
654 units
Distribution
Worldwide distribution
Product codes and lots
All serial numbers 1) AXIOM Luminos TF, Material #10093902 (no UDI) 2) AXIOM Iconos R100, Material #05895151 (no UDI) 3) AXIOM Iconos R200 C69, Material #10093961 (no UDI) 4) AXIOM Iconos R200 C20, Material #10093962 (no UDI) 5) AXIOM Luminos dRF, Material #10094200 (no UDI)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1000-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1000-2023Class IIOngoingVoluntary: Firm initiated
Z-1001-2023Class IIOngoingVoluntary: Firm initiated
Z-1002-2023Class IIOngoingVoluntary: Firm initiated