Z-1001-2023Class II
AXIOM Luminos Agile
Siemens Medical Solutions USA, Inc. / initiated 2022-12-09 / Open, Classified / root cause: manufacturing process
Siemens Medical Solutions USA, Inc began a recall of AXIOM Luminos Agile on . FDA classified it as Class II and gives the reason as "There is potential for the footrest to detach from the patient table during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1001-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Luminos Agile
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- There is potential for the footrest to detach from the patient table during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A Customer Safety Advisory Notification (CSAN) was distributed to affected consignees beginning December 9, 2022. The CSAN was distributed to inform the customers about a preventative measure to exclude any potential risk of patient injury caused by the footrest detaching from the patient table during examination. Together with the CSAN, an addendum to the operator manual is distributed containing additional details and illustrations on how to securely attach the footrest to further improve the understanding of handling the footrest. Refer to the CSAN with addendum and copies of the Instructions for Use for each reported system.
- Quantity in commerce
- 271 units
- Distribution
- Worldwide distribution
- Product codes and lots
- All serial numbers for AXIOM Luminos Agile, Material #10502200 (no UDI)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1000-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1001-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1002-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.