AirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0...
SunMed Holdings, LLC began a recall of AirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0 60105 1.5 60155 2.0 60205 3.0 60305 4.0 60405 on . FDA classified it as Class II and gives the reason as "There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1020-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Salter Labs
- Product
- AirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0 60105 1.5 60155 2.0 60205 3.0 60305 4.0 60405
- Product code
- CAE Airway, Oropharyngeal, Anesthesiology
- Reason for recall
- There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- On December 22, 2022, the firm notified affected customers through Urgent Medical Device Product Correction letters. Customers were informed that there is a discrepancy in the orogastric (OG) tube size indicated in the labeling in the instructions for use (IFU) versus the printed text on the device. Customers were provided with the correct recommended maximum OG tube/catheter sizes for each impacted Air-Q3G and Air-Q SPG model. If product has been further distributed, customers should forward the recall notice. If customers have questions about this recall notification, please contact Rob Yamashita, VP of Regulatory Affairs, at 616-259-8400, toll free at 800-433-2729, or [email protected].
- Quantity in commerce
- 43,290 (Eaches)
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.
- Product codes and lots
- ALL LOTS Size REF UDI/DI 0.0 60005 Each: 00814954020263 Box: 10814954020260 Case: 20814954020267 0.5 60055 Each: 00814954020270 Box: 10814954020277 Case: 20814954020274 1.0 60105 Each: 00814954020287 Box: 10814954021083 Case: 20814954020281 1.5 60155 Each: 00814954020294 Box: 10814954021090 Case: 20814954020298 2.0 60205 Each: 00814954020300 Box: 10814954021106 Case: 20814954020304 3.0 60305 Each: 00814954020317 Box: 10814954021113 Case: 20814954020311 4.0 60405 Each: 00814954020324 Box: 10814954021120 Case: 20814954020328
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1019-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1020-2023 | Class II | Ongoing | Voluntary: Firm initiated |