Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841...
Medtronic Perfusion Systems began a recall of Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass on . FDA classified it as Class II and gives the reason as "An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1026-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass.
- Product code
- DTZ Oxygenator, Cardiopulmonary Bypass
- Reason for recall
- An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
- FDA root cause
- Process change control
- Action
- Medtronic initially notified consignees on 12/12/2022 and followed up with an updated notification on 12/21/2022. Consignees were advised of two options: 1. Continue to use the Affinity Fusion Oxygenator without using the TMA (use other conventional perfusion circuit temperature monitoring methods); or 2. Continue to use the Affinity Fusion Oxygenator and the TMA for arterial temperature monitoring. Ensure there is minimal torque applied to the oxygenator TMA when attaching or detaching the Temperature Probe (Product Number ATP210). Also, minimize manipulation of the TMA-probe connection during the clinical procedure. Consignees were also advised to complete and return the Customer Confirmation Form, and notify personnel or facilities were affected units are located. On about 02/10/2023, Medtronic initiated an expansion and notified customers. Instructions in the customer letter remained the same.
- Quantity in commerce
- 4,462 units
- Distribution
- Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Honduras, Hungary, Iceland, Indonesia, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, Turkmenistan, Uganda, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Yemen.
- Product codes and lots
- Model CB841; Serial Number Range: 8111483548 - 8113999999; GTIN Number: 00763000225452 (Lot Numbers: 13394614, 13395228, 223819363, 223952969, 224315522, 224448403, 224772847, 224772848, 224772849); GTIN Number: 00763000225469 (Lot Numbers: 13392615, 13394868, 223476442, 223755429, 223796291, 223834097, 223834098, 224174020, 224188890, 224365018, 224419842, 224816179, 225222080, 225273494, 225273495, 225331432, 225332504, 225332505, 225332506, 225332508, 225661147)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1023-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1024-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1025-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1026-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1027-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1028-2023 | Class II | Ongoing | Voluntary: Firm initiated |