Z-1032-2023Class II
Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case
MEDLINE INDUSTRIES, LP - Northfield / initiated 2022-11-17 / Open, Classified / root cause: design
MEDLINE INDUSTRIES, LP - Northfield began a recall of Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case on . FDA classified it as Class II and gives the reason as "Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1032-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Non-sterile procedural trays labeled as MAJOR/MINOR, 3 per case
- Product code
- LRO General Surgery Tray
- Reason for recall
- Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Medline Industries, LP notified consignees via email and first-class mail on 11/21/2022. Consignees were instructed to over-label the sterile kits with a warning label indicating Warning To Avoid Injury, Please Handle Blades with Care . For the Non-Sterile kits, they are to over label the affected kits indicating to remove the blades and replace from a different sterile supply. No product is to be returned. Customers were also instructed to complete and return the response form.
- Quantity in commerce
- 160 cases (480 units)
- Distribution
- US Nationwide distribution in the states of AR, MD, WA, WI.
- Product codes and lots
- Item Number: DYNJ900416M, UDI/GTIN Case: 40195327031672, UDI/GTIN Each: 10195327031671, Lot Number: 22IBQ980
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1029-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1030-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1031-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1032-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1033-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1034-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1035-2023 | Class II | Ongoing | Voluntary: Firm initiated |