Z-1187-2023Class II

Empowr Vvc Knee Tibial Insert, Size 6, 22Mm, Ref: 346-22-706; Empowr Vvc Knee Tibial Insert, Size 7, 22Mm...

Encore Medical, LP / initiated 2022-12-22 / Open, Classified / root cause: labeling and UDI

Encore Medical, LP began a recall of EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE... on . FDA classified it as Class II and gives the reason as "Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1187-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
Empowr Vvc Knee Tibial Insert, Size 6, 22Mm, Ref: 346-22-706; Empowr Vvc Knee Tibial Insert, Size 7, 22Mm, Ref: 346-22-707; Empowr Vvc Knee Tibial Insert, Size 8, 22Mm, Ref: 346-22-708; Empowr Vvc Knee Tibial Insert, Size 9, 22Mm, Ref: 346-22-709; Empowr Vvc Knee Tibial Insert, Size 10, 22Mm, Ref: 346-22-710; Empowr Vvc Knee Tibial Insert, Size 11, 22Mm, Ref: 346-22-711; Empowr Vvc Knee Tibial Insert, Size 5 Bridge Up, 22Mm, Ref: 346-22-755; Empowr Vvc Knee Tibial Insert, Size 6 Bridge Down, 22Mm, Ref: 346-22-766; Empowr Vvc Knee Tibial Insert, Size 2, 25Mm, Ref: 346-25-702; Empowr Vvc Knee Tibial Insert, Size 3, 25Mm, Ref: 346-25-703; Empowr Vvc Knee Tibial Insert, Size 4, 25Mm, Ref: 346-25-704; Empowr Vvc Knee Tibial Insert, Size 5, 25Mm, Ref: 346-25-705; Empowr Vvc Knee Tibial Insert, Size 5 Bridge Up, 25Mm, Ref: 346-25-755; Empowr Vvc Knee Tibial Insert, Size 6 Bridge Down, 25Mm, Ref: 346-25-766
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On 12/22/22 recall notices were emailed to surgical agents who were asked to do the following: 1) Pass on the notice to all those within their organization or to any organization where potentially affected products have been transferred. 2) Return affected products. 3) Complete and return the Consignee Reconciliation notice. The following number was provided: Customer Service, 1-800-456-8696 On 2/8/23 recall notices were emailed with an expanded scope.
Quantity in commerce
216
Distribution
US: MN, IL, IN, NY, FL
Product codes and lots
REF/UDI-DI/Lot: 346-22-706/ 00190446219378/484V1013, 484V1014, 484V1015, 484V1016, 484V1017, 484V1018; 346-22-707/ 00190446219385/485V1001, 485V1002, 485V1003, 485V1004; 346-22-708/ 00190446219392/486V1005, 486V1006, 486V1007, 486V1008; 346-22-709/ 00190446219408/487V1001, 487V1002, 487V1003, 487V1004; 346-22-710/ 00190446219415/488V1003, 488V1004, 488V1005, 488V1006; 346-22-711/ 00190446219422/489V1003, 489V1004, 489V1005, 489V1006; 346-22-755/ 00190446219439/490V1001, 490V1002, 490V1003, 490V1004; 346-22-766/ 00190446219477/491V1002, 491V1003, 491V1004; 346-25-702/ 00190446219507/492V1002, 492V1003, 492V1004, 492V1005; 346-25-703/ 00190446219569/493V1001, 493V1002, 493V1003, 493V1004; 346-25-704/ 00190446219583/494V1001, 494V1002, 494V1003, 494V1004; 346-25-705/ 00190446219590/495V1001, 495V1002, 495V1003, 495V1004; 346-25-755/ 00190446219781/ 502V1001, 502V1002, 502V1003; 346-25-766/ 00190446219798/503V1003, 503V1004, 503V1005, 503V1006, 503V1007;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1187-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1187-2023Class IIOngoingVoluntary: Firm initiated