Cardinal Health Presource Packs IV START KIT-Intended for IV Access CATALOG #: 01A4553E
Cardinal Health 200, LLC began a recall of Cardinal Health Presource Packs IV START KIT-Intended for IV Access CATALOG #: 01A4553E on . FDA classified it as Class II and gives the reason as "Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1159-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Cardinal Health Presource Packs IV START KIT-Intended for IV Access CATALOG #: 01A4553E
- Product code
- LRS I.V. Start Kit
- Reason for recall
- Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Cardinal Health issued Urgent: Product Correction letter and Acknowledgement Form was sent overnight on November 18, 2022 to the direct consignees. Letter states reason for recall, health risk and action to take: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. We sincerely apologize for any inconvenience this notice may have caused you and your staff. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332
- Quantity in commerce
- 31,300 EA
- Distribution
- US Nationwide distribution in the states of GA, PA.
- Product codes and lots
- UDI-DI: 50195594389336(ea) 10195594389338(cs) Lot Number: 818636 Exp. Date: 3/1/2024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1157-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1158-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1159-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.